• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67PFPS45
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during pre-test, the device was not inflating fully.No patient injury was reported.
 
Additional Manufacturer Narrative
Two samples were received in used conditions.No damage was detected in visual inspection.When measuring the cuffs, the percentage for the symmetry was for sample 1: 33.33% and for sample 2: 37.83% these results are over 33.3% that is the minimum acceptable.Based on the test, the units are within spec.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15280030
Report Number3012307300-2022-15948
Device Sequence Number7166828
Product Code JOH
UDI-Device Identifier10351688518552
UDI-Public10351688518552
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2023
Device Model Number67PFPS45
Device Catalogue Number67PFPS45
Device Lot Number3708636
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Initial Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-