| Model Number |
67PFPS45 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during pre-test, the device was not inflating fully.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Two samples were received in used conditions.No damage was detected in visual inspection.When measuring the cuffs, the percentage for the symmetry was for sample 1: 33.33% and for sample 2: 37.83% these results are over 33.3% that is the minimum acceptable.Based on the test, the units are within spec.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Search Alerts/Recalls
|