It was reported that, during a thr surgery, the tip of a r3 lnr mpactor hd ii 36mm broke when impacting the liner.All broken pieces were removed from body and liner was set appropriately.Surgery was finished without any delay.No harm to the patient or further complications reported.
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirms the device broke into two pieces.Both pieces were returned for evaluation.The device shows signs of significant wear and use.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that during a thr surgery, the tip of a r3 lnr mpactor hd ii 36mm broke when impacting the liner.All broken pieces were removed from body and liner was set appropriately.Surgery was finished without any delay.No harm to the patient or further complications reported.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's condition is reported as unharmed.No patient impact is anticipated, based on the limited information provided.Should any additional clinical information be provided, this complaint will be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it cannot longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|