ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804275-18
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.User facility medwatch report: mw5111021.
 
Event or Problem Description
User facility medwatch report (mw5111021) received stating: "stent used during coronary ptca was unable to be advanced, so it was removed.There was difficulty removing it through the guard catheter.Everything was pulled out, catheter, guidewire and stent.Upon observation, the stent had become torn or damaged at the base.Abbott xience skypoint 2.75 mm x18 mm, ref 1804275-18, lot 1120641.Fda safety report id # (b)(4).Additional information received: after lesion preparation, the 2.75x18mm xience skypoint stent delivery system (sds) failed to advance due to heavy calcium and moderate tortuosity 90+% stenosed lesion in the mid right coronary artery (mrca).After removal of the sds, the distal end of the stent had a noted deformity due to the dottering technique on the calcium.Another 2.75mm x 18 mm xience skypoint stent was deployed.No additional information was provided.
 
Additional Manufacturer Narrative
A visual and dimensional inspection was performed on the returned device.The reported material split, cut or torn could not be confirmed.The first three rings at the proximal end of the stent implant were mangle.The stent implant was not torn as reported.The reported failure to advance could not be tested as it was based on operational context.The reported difficult to remove could not be tested due to the device condition.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported failure to advance and difficult to remove appear to be related to the operational context of the procedure; however, the investigation was unable to determine a conclusive cause for the reported material split/torn/cut.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15280205
Report Number2024168-2022-09081
Device Sequence Number11361093
Product Code NIQ
UDI-Device Identifier08717648233166
UDI-Public08717648233166
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/12/2023
Device Model Number1804275-18
Device Catalogue Number1804275-18
Device Lot Number1120641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer09/08/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-