STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5532-G-313
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that while resurfacing the patient's native patella, the surgeon decided to revise a 3x13 ps insert to a 3x14 ps insert.
 
Additional Manufacturer Narrative
Reported event: an event regarding revision involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: the reported device was not returned however photographs were provided for review.The photos show an explanted insert device without its locking wire.Damage is visible on the device.The damage is consistent with explantation process.Yellow discoloration is also observed on the tibial insert, the discoloration is consistent with the absorption of synovial fluid by the device.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient's knee was revised to resurface the patella which was not done at the index operation.During the surgery, the tibial insert was revised a 3x13 ps insert to a 3x14 ps insert.Reason for insert revision was not reported.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that while resurfacing the patient's native patella, the surgeon decided to revise a 3x13 ps insert to a 3x14 ps insert.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key15280286
Report Number0002249697-2022-01221
Device Sequence Number15245448
Product Code MBH
UDI-Device Identifier07613327050653
UDI-Public07613327050653
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Model Number5532-G-313
Device Catalogue Number5532-G-313
Device Lot NumberMMHK5P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/23/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight106 KG
-
-