Model Number 67SP055
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual inspection: visual inspection was performed at 12" under normal lighting to received unit, in order to detect any damage on the cuff, airline or pilot balloon.No damaged could be detected.Functional test: sample was filled with 5cc of air in order to see if there was any functional problem.Results: when inflating the unit with air, it was seen that cuff inflated but a part was not uniformly.The cuff was manipulated and the whole cuff inflated.However, extremely force was needed to inflate the cuff.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Event or Problem Description
It was reported that the cuff was sticking to the shaft of the device.No patient injury reported.
 
Additional Manufacturer Narrative
This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147., corrected data: corrected information provided in d2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15280307
Report Number3012307300-2022-15961
Device Sequence Number12554507
Product Code JOH
UDI-Device Identifier15021312000757
UDI-Public15021312000757
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67SP055
Device Catalogue Number67SP055
Device Lot Number2884258
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/10/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1