MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 3549112
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Seroma (2069)
Date of Event 06/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation summary: a manufacturing record evaluation was performed for the finished device (b)(4) number, and no non-conformances were identified.Upon visual evaluation of the image provided in the complaint, no apparent damage or visual anomalies were observed.A seroma is a build-up of fluid around the implant.Symptoms from a seroma may include swelling, pain, and bruising.A seroma may occur soon after surgery, or any time later on, which would be referred to as a late seroma.While the body may absorb seromas, some may require surgery for drainage.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: seroma mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient who underwent breast augmentation with a 350cc mentor cpx 4 breast tissue expander with suture tabs experienced early onset seroma post procedure.As a result, explant was performed on (b)(6) 2022.
 
Additional Manufacturer Narrative
Device evaluation summary: the investigation of the returned photograph was completed by the failure analysis lab on september 23, 2022.Upon visual evaluation of the image provided in the complaint, no apparent damage or visual anomalies were observed.The investigation of the returned device was completed by the failure analysis lab on september 28, 2022.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned tissue expander.During the visual evaluation, no apparent damage or visual anomalies were observed on the tissue expander.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The mentor failure analysis lab received the device for evaluation on september 2, 2022.The analysis has begun but is not complete at this time.When the investigation and analysis have been completed, a supplemental report will be submitted.The procedure was a breast augmentation revision.Replacement with a 350cc mentor cpx 4 breast tissue expander with suture tabs was performed with explant.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key15280373
Report Number1645337-2022-09900
Device Sequence Number7308009
Product Code LCJ
UDI-Device Identifier00081317006664
UDI-Public00081317006664
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K130813
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/08/2025
Device Model Number3549112
Device Catalogue Number3549112
Device Lot Number9666499
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/02/2022
09/23/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date09/22/2022
09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-