• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number ACU0T0
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
A health professional reported with a description of faulty intraocular lens (iol) additional information has been received and stated that, there was a problem with the release of the lenses.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The device with the lens was returned.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger is at the trailing optic edge.The lens is advanced to mid-nozzle.The trailing haptic is folded in on the optic.The leading haptic is extended.No problems are observed with the returned device.The plunger, lens and haptic positions are acceptable.The leading haptic was straight.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the dfu (directions for use).A non-qualified viscoelastic was indicated a straight leading haptic may occur: ¿ due to the normal folding variations as indicated in the dfu diagrams.¿ if the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.¿ if there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A non-qualified viscoelastic was indicated.The root cause may be related to a failure to follow the ifu (instructions for use).A non-qualified viscoelastic was used in the device.The ifu instructs: during device preparation and implantation of the company iq iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Lens may become stuck in the device: ¿ if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to become ¿stuck¿ in the device ¿ due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The product has not been received to evaluate.File will be reopened when product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15281308
Report Number1119421-2022-01837
Device Sequence Number7307521
Product Code HQL
UDI-Device Identifier00380652395359
UDI-Public00380652395359
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2024
Device Model NumberACU0T0
Device Lot Number15295356
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer10/03/2022
11/19/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date10/27/2022
12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEALON VISCOELASTIC
Patient SexUnknown
-
-