MEDTRONIC MEXICO CUSTOM TUBING PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BB11S45R4
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: upon receipt in our quality analysis lab, visual inspection showed no outward signs of any damage.The device had a test mark, which confirmed the vrv was tested prior to distribution.Functional testing of the returned device was performed at 0.5 l/min.The instructions for use provided an approximation for first time use of negative and positive pressures to open the umbrella relief valves at 0.5 l/min blood flow are approximately -205 mmhg and 318 mmhg.The results indicated the negative pressure umbrella opened at -178 mmhg and the positive pressure umbrella valve opened at 324 mmhg verified the functionality for the valves for the returned device.The acceptable function of the umbrella valves also confirmed that the duck bill valve functioned as intended.The device appeared to operate as expected with no evidence of leaking.Conclusion: the complaint is unconfirmed for leaking valve and there was no observed damage to the device and performance testing indicated that the device functions as intended.After evaluation it was determined there was no product failure that occurred during the reported event.The device history record was not reviewed as the returned product analysis found no evidence of manufacturing issues with the returned device.Assessment against the medtronic risk management file dfmeca document indicates that the current risk zone does not exceed the risk zone predicted in the product dfmeca; therefore, no capa will be initiated at this time.There were no adverse patient effects as a result of this incidence.Review of the complaint file indicates sufficient information was provided for completion of this investigation.Medtronic will continue to monitor for future occurrences and trends per trend analysis procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during use of a custom tubing pack, the customer reported that while giving cardioplegia, the one way valve leaked.The device was replaced to complete the procedure.There was no patient impact associated with this event.Medtronic received additional information that the leak originated from the one way valve.As this is a vent line, air was seen in line.Blood was seen on the floor and about 20cc of blood leaked, but no transfusion was required.No damage noted on the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM TUBING PACK
Common Device Name
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15281352
Report Number9612164-2022-03158
Device Sequence Number7054725
Product Code DWE
UDI-Device Identifier00763000640088
UDI-Public00763000640088
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K171979
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2024
Device Model NumberBB11S45R4
Device Catalogue NumberBB11S45R4
Device Lot Number224019307
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2022
Initial Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-