• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Model Number VERSION 1.0
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a recanalization procedure, the helix allegedly broken.It was further reported that the part was stuck in the patient.Reportedly, broken wire was retrieved with endovascular snare.The procedure was successfully completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device not returned.
 
Additional Manufacturer Narrative
The medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was not returned for evaluation.Therefore, the investigation is inconclusive for the reported break issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a recanalization procedure, the helix allegedly broken.It was further reported that the part was stuck in the patient.Reportedly, broken wire was retrieved with endovascular snare.The procedure was successfully completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Common Device Name
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15281399
Report Number3008439199-2022-00108
Device Sequence Number2291800
Product Code MCW
UDI-Device Identifier07640142810384
UDI-Public(01)07640142810384
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K211738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERSION 1.0
Device Catalogue Number80207
Device Lot Number220019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-