MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DESIGN OPTIONS®; EPIDURAL ANESTHESIA KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DESIGN OPTIONS®; EPIDURAL ANESTHESIA KIT Back to Search Results
Model Number 555110
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported by the user facility: detailed inquiry description: the patient required surgical intervention to remove the intact epidural.The neurosurgeon found a defect in the catheter sheath on removal.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Three photos were provided by the reported facility for evaluation.Visual evaluation of the photos showed lidstock packaging confirming the reported item and lot# and a partial piece of a used catheter that was sheared and knotted.Per the manufacturers investigation this defect is not likely to have happened during the manufacturing process.It is believed to happen during application.User will be referred to the ifu which states, "precaution: advancing the catheter more than 5 cm past the needle tip may increase the likelihood of kinking or knotting.Caution: do not withdraw catheter through the epidural needle due to the possibility of shearing or kinking the catheter.When removing the catheter, excessive force should never be applied.If the catheter becomes difficult to withdraw, consult standard textbooks for specific techniques." review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESIGN OPTIONS®
Common Device Name
EPIDURAL ANESTHESIA KIT
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15281455
Report Number2523676-2022-00427
Device Sequence Number2467796
Product Code OGE
UDI-Device Identifier04046964313653
UDI-Public(01)04046964313653
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number555110
Device Catalogue Number555110
Device Lot Number0061817143
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/17/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-