• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FB INSRT SZ 4 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FB INSRT SZ 4 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-40-407
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); Pain (1994); Synovitis (2094); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 11/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
A call from a customer was received stating that "i want to know who i can talk about my knee replacement.I'm having a problem with it and it's been recalled even before my doctor put it in me.And since then i start to have problem with it.Should i call a lawyer for the compensation?" doi: (b)(6) 2019; dor: (b)(6) 2020; unknown knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).(b)(4) medical records ad 10 aug 2022 were reviewed on 08/22/2022 by clinician.On (b)(6) 2019 the patient had a left total knee arthroplasty to address degenerative joint disease.Depuy components, including depuy patella were used in conjunction with depuy cement x2.(part/lot page 4,5 of 848).On (b)(6) 2020, the patient had a revision left total knee to address instability and aseptic loosening.Prior to surgery the patient had pain, swelling, effusions, stiffness, and limping.During the procedure the surgeon observed significant synovitis, osteolysis.The tibial tray, insert, and femoral component were revised.In the operative notes, there were no specific loosening components noted, nor any interfaces provided.Competitor components were implanted during this surgery.(b)(6) 2021 right partial knee arthroplasty ¿ competitor products were used.Doi: (b)(6) 2019 dor: (b)(6) 2020 (left knee).Patient medical history: mild asthma, obesity, hypothyroidism, gerd, history of gastric bypass surgery.Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Medical records received.(b)(6) 2021: patient reports left knee pain at night.Treated with po pain meds.(b)(6) 2021: patient reports her left knee feels warm and stiff.X-ray reported no issues.No treatment was indicated and surgeon will continue to observe left knee.Only depuy implant is the patella (1518-20-035) which was implanted on (b)(6) 2019.All other implants are competitor.This was captured in (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a4, b5 and b7.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
The medical records received were for a cardiac stress test and echocardiogram.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Medical records received: update: medical records received on ad 4 mar 2024 were reviewed by a clinician to identify patient harms/product issues.The attached medical records reviewed clinic and pt notes dated (b)(6) 2022 due to diffuse, systemic joint pain and swelling attributed to the patient's fibromyalgia.No treatments were provided.The joint pain is already captured on (b)(4).There is no new information that would impact existing coding and/or us fda mdr determinations.Mrr activity closed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Is report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Medical records included a multitude of various medical procedures and clinic visits that are either duplicative or not relevant to the event that is captured on this pc.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : a call from a customer was received stating that "i want to know who i can talk about my knee replacement.I'm having a problem with it and it's been recalled even before my doctor put it in me.And since then i start to have problem with it.Should i call a lawyer for the compensation?" the product was not returned to depuy synthes, however photos were provided for review.(b)(4), (b)(6) 2023' (b)(4),(b)(6) 2023' (b)(4), (b)(6) 2024.All available x-rays were reviewed, and no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was unconfirmed as the observed condition of the [attune ps fb insrt sz 4 7mm] would not contribute to the reported adverse event.  based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the device lot number is unknown, therefore a device history review could not be performed. if the lot/serial number becomes available, the record will be re-assessed.
 
Event or Problem Description
The attached medical records were for a bladder sling procedure to treat incontinence and a carpal tunnel release procedure.Pt notes dated (b)(6) 2019 & 2020: patient received physical therapy to treat pain, swelling, walking difficulty, and a ¿clicking, popping, and grinding¿ sensation in left knee extensor mechanism.Physical therapist notes reduced rom in the left knee with stiffness.Pt also identified patellar clunk during passive rom.Patient continued to improve through the sessions.The patellar clunk and extensor mechanism sensations are new findings on medical record dated (b)(6) 2024.Clinic note dated (b)(6) 2020: patient reports continued pain along the anterior portion of the left knee.The left knee patellar clunk and extensor mechanism sensations are new findings on medical record dated (b)(6) 2024.The attached medical records are related to the patient¿s foot and ankle procedures dated 2016-2017.These are unrelated to the left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 4 7MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15281841
Report Number1818910-2022-16490
Device Sequence Number19402943
Product Code NJL
UDI-Device Identifier10603295049807
UDI-Public10603295049807
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-40-407
Device Catalogue Number151640407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer11/27/2023
01/17/2024
01/26/2024
02/22/2024
03/04/2024
03/28/2024
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date12/06/2023
01/24/2024
02/04/2024
03/04/2024
03/08/2024
03/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 4 CEM.; ATTUNE MEDIAL DOME PAT 35MM.; ATTUNE PS FB INSRT SZ 4 7MM.; ATTUNE PS FEM LT SZ 4 CEM.; SMARTSET HV BONE CEMENT 40G.; SMARTSET HV BONE CEMENT 40G.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
Patient Weight93 KG
-
-