| Model Number |
0684-00-0480-01C |
| Medical Device Problem Codes |
Difficult to Insert (1316); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event site postal code: (b)(6).Event site telephone# (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that there was resistance while inserting the intra-aortic balloon (iab) and blood was found in the iab.They then pulled out the iab and inserted a new one without further issues.There was no patient harm or adverse event reported.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter.A visual examination of the product detected the inner lumen within the membrane was completely separated within a kink approximately 27.7cm from iab tip.A laboratory insertion test was unable to be performed due to the catheter¿s returned condition.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no other leaks were detected.The break found in the inner lumen appears to have been the result of a severe kink, which eventually failed causing the reported problems.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
| |
|