| Model Number |
407360 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure, a perforation in the sheath was noted during removal from the patient at the femoral site.Blockage of the needle at the level of the iliac vein was noted resulting in a need to replace the sheath.There were no adverse patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8f swartz braided introducer sheath and dilator were received for evaluation.The sheath and dilator had been perforated proximal to the distal tip.A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The brk needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ the cause of the sheath and dilator perforation is consistent with not following the instructions for use.
|
| |
|