MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 45, 63 CM LENGTH, 8F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 45, 63 CM LENGTH, 8F; INTRODUCER, CATHETER Back to Search Results
Model Number 407360
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the procedure, a perforation in the sheath was noted during removal from the patient at the femoral site.Blockage of the needle at the level of the iliac vein was noted resulting in a need to replace the sheath.There were no adverse patient consequences.
 
Additional Manufacturer Narrative
One 8f swartz braided introducer sheath and dilator were received for evaluation.The sheath and dilator had been perforated proximal to the distal tip.A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The brk needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ the cause of the sheath and dilator perforation is consistent with not following the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 45, 63 CM LENGTH, 8F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15282305
Report Number3005334138-2022-00482
Device Sequence Number7054215
Product Code DYB
UDI-Device Identifier05414734205559
UDI-Public05414734205559
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K052644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model Number407360
Device Catalogue Number407360
Device Lot Number8531710
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-