MERIT MEDICAL SYSTEMS, INC. PERFORMA®; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 00884450002605
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during a heart catheterization with angioplasty on a pediatric patient, the catheter cracked while inside the patient.No patient injury to report.
 
Additional Manufacturer Narrative
The suspect medical device has been returned for evaluation.The device was examined visually and microscopically and no kinks or cracks were observed.The complaint could not be confirmed.No kinks or cracks were observed along the shaft, around the marker bands or in the tip section.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMA®
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC. - PEARLAND
14646 kirby drive
houston TX 77047
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key15283159
Report Number3010665433-2022-00088
Device Sequence Number19402811
Product Code DQO
UDI-Device Identifier00884450002605
UDI-Public00884450002605
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K000659
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00884450002605
Device Catalogue Number7742-20M100SH
Device Lot NumberE2425457
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient SexUnknown
-
-