• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC 1.6MM X 5MM SCREW; SCREW, FIXATION, INTRAOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC 1.6MM X 5MM SCREW; SCREW, FIXATION, INTRAOSSEOUS Back to Search Results
Model Number 208-1605
Medical Device Problem Code Signal Artifact/Noise (1036)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Attempts to obtain additional information regarding this event are being made.The customer was notified per the instructions for use (ifu): the devices have not been tested for safety and compatibility in the mr environment, radiofrequency (rf) heating, migration due to magnetically induced displacement, or image artifact in the mr environment, and scanning a patient who has these devices may result in patient injury.The investigation is currently pending.A follow up report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported the patient underwent a procedure on (b)(6) 2022 to remove a brain tumor and the osteomed devices were used to close the craniotomy.It was later reported by the surgeon (event date unknown) that the surgeon is unable to plan the sequential radiotherapy treatment for the patient's cancer due to the artifacts generated by the titanium plates and screws on "the mdr exam".It is currently unknown what kind of imaging was performed.It was reported the devices "were not removed from the patient yet".No adverse patient conseqiences were reported.This report is related to report numbers 2027754-2022-00040, 2027754-2022-00041, 2027754-2022-00042, 2027754-2022-00043, 2027754-2022-00044, and 2027754-2022-00045 for the other implanted devices involved in this event.
 
Additional Manufacturer Narrative
The devices were not received for evaluation.Four (4) part numbers and lot numbers were reported.All four dhrs were reviewed.Lot 1145443 (p/n 208-1605) packaged five hundred seventy-five (575) parts from bulk lot 1144717 one screw per bag on 1/20/2020.Lot 1145443 was final released to stock on 1/21/2020.All inspections passed and no issues were identified in the dhr review.The bulk lot 1144717 was machined from raw material lot 1143186, and the dhr for lot 1143186 was reviewed as well.The ti 6al-4v (mrtb0002) from was purchased from the vendor and was received on 10/30/2019.All material certifications conform to astm f136-13, the standard specification for wrought titanium-6aluminum-4vanadium eli (extra low interstitial) alloy for surgical implant applications.All inspections passed and no issues were identified in the dhr review.Lot 1062827 (p/n 213-0014) manufactured and packaged one hundred sixty (160) parts released to stock on 1/7/2014.The lot was machined from raw material mrts0006 (0.23 x 5 x 8 gr-2 cp ti sheet), lot 1056955.All inspections passed and no issues were identified in the dhr review.The dhr for the titanium from lot 1056955 (mrts0006) was reviewed as well.The lot was purchased from the vendor and was received on 4/1/2013.All materials certifications conformed to astm f-67, the standard specification for unalloyed titanium, for surgical implant applications.All inspections passed and no issues were identified in the dhr review.Lot 1161156 (p/n 208-1604) packaged fifty-five (55) parts from bulk lot 1157552 one screw per bag on 8/4/2021.All inspections passed and no issues were identified in the dhr review.Bulk lot 1157552 was machined from the raw material lot 1157696 and the dhr for the raw material (mrtb0002) was reviewed as well.The ti 6al-4v was purchased from the vendor and was received on 4/16/2021.All material certifications conform to astm f136-13, the standard specification for wrought titanium-6aluminum-4vanadium eli (extra low interstitial) alloy for surgical implant applications.All inspections passed and no issues were identified in the dhr review.Lot 1157318 (p/n 204-1203) packaged two hundred forty (240) parts one screw per bag on 5/6/2021.The packaging lot used two hundred one (201) screws from lot 1149595 and thirty-nine (39) screws from lot 1155997.All inspections passed and no issues were identified in the dhr review.Bulk lot 1149595 was machined from raw material lot 1147319 (mrtb0002) and the dhr was reviewed for the raw material as well.The lot was purchased from the vendor and was received on 4/1/2020.All material certifications conform to astm f136-13, the standard specification for wrought titanium-6aluminum-4vanadium eli (extra low interstitial) alloy for surgical implant applications.All inspections passed and no issues were identified in the dhr review.Bulk lot 1155997 was machined from the raw material lot 1154238, mrtb0002 and the dhr for the raw material was reviewed as well.All material certifications conform to astm f136-13, the standard specification for wrought titanium-6aluminum-4vanadium eli (extra low interstitial) alloy for surgical implant applications.All inspections passed and no issues were identified in the dhr review.Magnetic fields can be very fast when near metal, which may lead to dephasing and signal loss.The lack of signal may cause a black metal artifact to appear on the image.The ifus for the products used all include a section titled "mri status".This section warns that the products were not tested for safety or compatibility in the mr environment.Furthermore, the literature states that the products were not tested for image artifacts in the mr environment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.6MM X 5MM SCREW
Common Device Name
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key15283238
Report Number2027754-2022-00039
Device Sequence Number14026273
Product Code DZL
UDI-Device Identifier00845694000956
UDI-Public(01)00845694000956(10)114543(30)1(11)200117
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number208-1605
Device Catalogue Number208-1605
Device Lot Number1145443
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer08/28/2022
09/12/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date08/29/2022
09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
Patient SexMale
-
-