Model Number 1294-16-515
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 08/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient presented with pain and some feelings of mechanical failure due to prior tka.Upon surgery, the surgeon found tracking abnormally and knee balance loose in flexion and extension.Patella and tibial poly bearing were replaced.No information/paperwork available.The patella was damaged upon removal.The cement manufacturer was not indicated.No loosening and no surgical delay.Doi: (b)(6) 2012; dor: (b)(6) 2022; affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMP RPS INS STD+ 15MM
Common Device Name
LCS COMPLETE : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15283565
Report Number1818910-2022-16548
Device Sequence Number20624634
Product Code NJL
UDI-Device Identifier10603295023265
UDI-Public10603295023265
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1294-16-515
Device Catalogue Number129416515
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer09/05/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK KNEE PATELLA LCS
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-