Model Number 45-751019
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.If the device or additional information is received, microvention, inc., will submit a supplemental report.The instructions for use (ifu) identifies premature coil detachment as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that the first coil used started to unravel/stretch during pulling for repositioning.The coil was removed from the patient.The second coil deployed detached unintentionally during pulling for repositioning and coil got stuck in the microcatheter when attempted to manually push out.The entire coil was removed together with the microcatheter.Procedure outcome was successful and patient was stable.
 
Additional Manufacturer Narrative
Summary device evaluation: only the implant was returned for evaluation.Visual inspection of the implant found the proximal end to be severely stretched.The physical evaluation of the device could not determine the exact circumstances that led to this condition, but the stretching is indicative of the coil encountering resistance followed by retraction forces that exceeded the strength of the coil winding.The pusher was not returned for evaluation, so the investigation could not definitively determine the mode of separation (i.E.Thermal or tensile), nor could the physical evaluation determine when or where the separation occurred.The microcatheter used in the procedure was not returned, so the investigation could not determine if a condition existed that would have caused or contributed to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR CX 35 DETACHABLE
Common Device Name
VASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key15283715
Report Number2032493-2022-00323
Device Sequence Number2504322
Product Code KRD
UDI-Device Identifier00812636021072
UDI-Public(01)00812636021072(11)211202(17)261130(10)0000131359
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K151358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number45-751019
Device Catalogue Number45-751019
Device Lot Number0000131359
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-