• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-EX-125MIC145
Medical Device Problem Codes Mechanical Problem (1384); Unintended System Motion (1430)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 07/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During the second treatment of the dorsalis pedis (dp) artery with the diamondback exchangeable orbital atherectomy device (oad), the crown jumped backwards and stopped spinning.The oad was removed, and a perforation of the dp artery was observed.Balloon angioplasty was attempted to resolve the perforation.A non-csi balloon was unable to cross the lesion and became stuck on the viperwire advance guide wire.Another attempt was made with a different non-csi balloon, however, the issue persisted.The wire and balloon were removed and a kink proximal of the spring tip was observed on the wire.No further intervention was performed to resolve the perforation.The patient was stable, and the perforation will continue to be monitored.
 
Additional Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.Related manufacturer report number: 3004742232-2022-00197.Csi id: (b)(4).
 
Additional Manufacturer Narrative
The oad was returned to csi for analysis.During the investigation, device stall events were observed.Further analysis revealed adhered biological mater on the driveshaft crown which prevented a test wire from passing through.Analysis did not identify any damage that may have contributed to the accumulating material.The morphology and exact root cause of the accumulated biological material was unable to be determined.The reported event of crown jumping and perforation were unable to be confirmed.When tested, the device functioned as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key15284177
Report Number3004742232-2022-00196
Device Sequence Number7111096
Product Code MCW
UDI-Device Identifier10852528005817
UDI-Public(01)10852528005817(17)230430(10)375988
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDBP-EX-125MIC145
Device Catalogue Number7-10030-01
Device Lot Number375988
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-