ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
SCM-520 |
| Medical Device Problem Codes |
Malposition of Device (2616); Migration (4003)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side "bottoming out and malposition, worse on the left" and "large air pocket/crease in implant." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: weight to the spec, opening, crease fold.A microscopic analysis was performed which identify a striated edge opening.Based on the device analysis the final assessment is: a striated edge opening on posterior side assessed as surgical damage.A review of the device history record has been completed.No deviations or non-conformances noted.The reported events are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: "bottoming out" of the device.
|
| |
|
Search Alerts/Recalls
|
|
|