• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HERCULES III; UNIVERSAL STABILIZER ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HERCULES III; UNIVERSAL STABILIZER ARM Back to Search Results
Model Number T401161
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported to terumo cardiovascular a beating heart issue.It is still unknown when the event occurred, if it resulted in a delay in the procedure, if the product was changed out or if the surgery was completed successfully, if it caused or contributed to an injury to the patient or if there was a blood loss.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.
 
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on august 24, 2022.Upon further investigation of the reported event, the following information is new and/or changed: g6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H6 (identification of evaluation codes 3331, 4114, 3221, 4315).Type of investigation #1: 3331 - analysis of production records.Type of investigation #2: 4114 - device not returned.Investigation findings: 3221 - no findings available.Investigation conclusions: 4315 - cause not established.The affected sample was not returned, and a detailed description of the issue was not provided; therefore, a thorough investigation could not be performed and a definitive root cause could not be determined.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event or Problem Description
Additional information indicates that the cable needs repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES III
Common Device Name
UNIVERSAL STABILIZER ARM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key15284542
Report Number1124841-2022-00145
Device Sequence Number11332417
Product Code MWS
UDI-Device Identifier00699753450493
UDI-Public(01)00699753450493
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2024
Device Model NumberT401161
Device Catalogue NumberN/A
Device Lot NumberT01932
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer10/13/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-