(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: ¿ when was the suture torn (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify ans : during handling prior to use on patient.¿ could you please confirm the quantity of sutures torn? ans : 4pc.¿ did the event occur during one or multiple patient procedures? ans : one.¿ what is the total number of procedures? information not available.¿ have any of these events been previously reported to ethicon? if so, provide the respective reference number(s).Information not available.¿ if this event occurred in multiple procedures, please provide information requested above for each patient event.Investigational summary: one opened box with twenty-seven packets of product code y359h were returned to ethicon for evaluation.Visual and functional test were performed on the retuned device.As per the sampling plan, a visual inspection was performed on thirteen samples, no defects were found on the packages.The packets were opened, the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no issue related to breakage suture or anomalies were observed during evaluation.Also, the functional test was performed using a instron and the tensile force were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|