Doctor indicated chipped tip, damaged.When removing the healing abutment, the driver tip was used.It was used for the first time, but the tip part was damaged with almost no torque applied.The operation has been postponed, but there is no health hazard.Tooth site # 29 (universal).
|
|
This report is being submitted to report additional information.The following sections are being reported: d4: unique identifier (udi) number; d9: device availability; h2: if follow-up, what type; h3: device evaluated by manufacturer; h4: device manufacturer date; h6: type of investigation; h6: investigation findings; h6: investigation conclusions; h10: additional narratives/data.The product was returned, and the reported event was confirmed following visual evaluation.Based on the evaluation, device malfunction has occurred.However, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported device that did or could cause or contribute to the reported event.Monthly post market trending review identified no adverse trends or actionable trends for the reported event or device.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product was within specifications and conforming when it left zimvie.Dhr review for the lot had revealed no deviations or non-conformance's which could have caused or contributed to the reported event.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
|