MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Results from the product history record review indicated the product met release criteria.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An other health professional reported that, during intraocular lens (iol) implant procedure, the lens was stuck in cartridge.There was patient contact, however no patient impact was involved.The procedure was completed with back up iol.Additional information was requested.
 
Additional Manufacturer Narrative
The product was returned for analysis and damage was observed to the iol.No cartridge was returned.Iol received adhered to a surgical swab.Solution was dried on both surfaces of the optic and haptics.The haptics are intact.The optic is cracked/fractured and scratched/marked-rejectable with fibers.The iol met specifications when dimensionally measured for edge thickness and plan view.No cartridge received.The product investigation could not identify the root cause for the reported complaint "lens stuck in cartridge" as only the iol was returned for evaluation.No cartridge was returned.Due to this, we are unable to complete a thorough investigation to determine a root cause.The dimensions of the lens were tested using an approved template and results show the lens dimensions to be within specification.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15285101
Report Number9612169-2022-00411
Device Sequence Number2278808
Product Code HQL
UDI-Device Identifier00380655093245
UDI-Public00380655093245
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.225
Device Lot Number25389067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer10/20/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC; UNKNOWN CARTRIDGE
Patient SexUnknown
-
-