| Model Number |
VICM5 13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm vicm5 13.2 implantable collamer lens of -9.0 diopter was implanted into the patients left eye (os) on (b)(6) 2022.On (b)(6) 2022 the lens was removed due to a high vault.A replacement lens was implanted of unknown model and size.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9-return date: 31-aug-2022.H3 - device evaluation: the lens was returned in liquid in a vial.Visual inspection found the optic and haptic torn.Dimensional inspection found the lens to be within device specifications.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5: the reporter indicated that a 13.2mm vicm5 13.2 implantable collamer lens of -8.50 diopter was implanted into the patients right eye (od) on (b)(6) 2022.On (b)(6) 2022 the lens was explanted due to a high vault.A replacement lens was implanted of unknown model and size on an unknown date.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5: the reporter indicated that a 13.2mm vicm5 13.2 implantable collamer lens of -9.00 diopter was implanted into the patients right eye (od) on (b)(6) 2022.On (b)(6) 2022 the lens was explanted due to a high vault.A replacement lens was implanted of unknown model and size on an unknown date.Claim# (b)(4).
|
| |
|