No device was returned but radiographs and a photo were provided confirming the device failure.There was no rep present in the case, and the hospital discarded the involved plate and screws.The patient did not experience any pain or adverse symptoms, their post op activity levels are unknown and fusion can not be confirmed.The root cause of the failure is unknown but review of the information provided suggests possible proud screw placement combined with excessive force during lock engagement or excessive post operative activity levels to be potential contributing factors.No additional investigation can be completed.Label review ".Potential adverse events and complications; as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications known to occur include: bending, fracture or loosening of implant component(s), loss of fixation." "the nuvasive acp system is designed to assist in providing an adequate biomechanical environment for fusion.If a delayed union or nonunion occurs the implant may fail due to metal fatigue.Patients should be fully informed of the risk of implant failure." ".Care should be taken to ensure that all components are ideally fixated prior to closure." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".Post-operative warnings: during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." (b)(4).Device disposed of at user facility.
|