| Model Number |
0624-149-55 |
| Medical Device Problem Codes |
Break (1069); Smoking (1585); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during preparation for a cervical conization procedure, this fiber was left on a cloth that had been placed on the patient.The console screen status remained in ready mode while the laser fiber was outside the patient.The cloth suddenly caught fire; the fire was quickly extinguished and there has no harm to the patient or medical staff.The procedure was completed with the same fiber.It is believed that the system was mistakenly activated, leading to the fire.Additional information was received indicating that the handpiece showed burn marks, indicating that the laser had been fired inside the handpiece.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 device code: change to reflect break.
|
| |
|
Event or Problem Description
|
|
It was reported that, during preparation for a cervical conization procedure, this fiber was left on a cloth that had been placed on the patient.The console screen status remained in ready mode while the laser fiber was outside the patient.The cloth suddenly caught fire; the fire was quickly extinguished and there was no harm to the patient or medical staff.The procedure was completed with the same fiber.It is believed that the system was mistakenly activated, leading to the fire.Additional information was received indicating that the handpiece showed burn marks, indicating that the laser had been fired inside the handpiece.This is indicative of a fiber break.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our quality assurance laboratory, this fiber underwent visual analysis.The returned fiber was in two pieces with the fiber body broken between the connector and the tip.The tip and distal end of the fiber body exhibited evidence of burning.Visual analysis confirmed the reported clinical observation.
|
| |
|
Event or Problem Description
|
|
It was reported that, during preparation for a cervical conization procedure, this fiber was left on a cloth that had been placed on the patient.The console screen status remained in ready mode while the laser fiber was outside the patient.The cloth suddenly caught fire; the fire was quickly extinguished and there has no harm to the patient or medical staff.The procedure was completed with the same fiber.It is believed that the system was mistakenly activated, leading to the fire.Additional information was received indicating that the handpiece showed burn marks, indicating that the laser had been fired inside the handpiece.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our quality assurance laboratory, this fiber underwent visual analysis.The returned fiber was in two pieces with the fiber body broken between the connector and the tip.The tip and distal end of the fiber body exhibited evidence of burning.Visual analysis confirmed the reported clinical observation.
|
| |
|
Event or Problem Description
|
|
It was reported that, during preparation for a cervical conization procedure, this fiber was left on a cloth that had been placed on the patient.The console screen status remained in ready mode while the laser fiber was outside the patient.The cloth suddenly caught fire; the fire was quickly extinguished and there has no harm to the patient or medical staff.The procedure was completed with the same fiber.It is believed that the system was mistakenly activated, leading to the fire.Additional information was received indicating that the handpiece showed burn marks, indicating that the laser had been fired inside the handpiece.No patient complications were reported.
|
| |
|