SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 22MM; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
04.211.022 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional procode: hrs.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Sterile part #: 04.211.016s sterile lot #: 9l03236 release to warehouse date: 31 jan 2022 manufacturing site: (b)(4) supplier: (b)(4) ag.Non-sterile part #: 04.211.016 non-sterile lot #: 620p367 manufacturing site: (b)(4) release to warehouse date: 18 jan 2022 supplier: synthes usa hq, inc.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022 during an osteosynthesis for right distal humerus fracture, four (4) consecutive locking screws failed to lock into the four holes of a plate.The surgeon had been following the surgical technique and had started inserting the locking screws using a drill sleeve.The surgeon consulted with the sales rep and ultimately decided to proceed using a new plate.The surgery was completed successfully with less than 30 minutes of delay.This report is for a 2.7mm ti va lckng scr slf-tpng with t8 stardrive recess 22mm.This is report 2 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: a device history record (dhr) review was conducted: sterile part: : 04.211.016s, sterile lot no : 9l03236 , release to warehouse date: 31 jan 2022, manufacturing site: werk selzach, supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Non-sterile part # 04.211.016, non-sterile lot # 620p367, manufacturing site: werk selzach, release to warehouse date: 18 jan 2022, supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|