MAUDE Adverse Event Report: CORCYM CANADA CORP. PERCEVAL SUTURELESS AORTIC HEART VALVE; TISSUE HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORCYM CANADA CORP. PERCEVAL SUTURELESS AORTIC HEART VALVE; TISSUE HEART VALVE Back to Search Results
Model Number PVS27
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 08/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer was informed that percival heart valve size 27 was implanted and explanted intra-operatively on (b)(6) 2022.Reportedly, the annulus diameter was 28.8mm on preoperative ct.During the sizing procedure, the xl white obturator was moderately resistant, so the xl size valve was implanted.However, after de-clamp, echo showed paravalvular leakage.After re-clamp, it was found that the part of the device has fallen to the lvot side.As such, device was explanted and epic 27mm was implanted instead.Based on the medical judgement received, the fall of the device was probably due to a larger native annulus.More than 20 minutes was added to the cross clamp time as a result of the reported event.Based on the information received, there was no concomitant procedure and echo is unavailable.There was no positioning difficulty, no abnormal geometry in the patient's annulus, and no problem with the device functionality.
 
Additional Manufacturer Narrative
Device was not returned from the site.
 
Additional Manufacturer Narrative
The manufacturing and material records for the perceval heart valve, model #icv1211 , s/n #(b)(6), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model #icv1211) perceval heart valve at the time of manufacture and release.
 
Additional Manufacturer Narrative
Updated sections b4, g3, g6, h2, h6, h11.The manufacturing and material records for the perceval heart valve, model #icv1211, s/n # (b)(6), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model #icv1211) perceval heart valve at the time of manufacture and release.Based on the available information, there was no positioning difficulty, no abnormal geometry in the patient's annulus, and no problem with the device functionality.Further, no deficits were noted during the production records review.Since device is not available for return as indicated by the site, no other investigation can be performed, and the root cause remains undetermined.Should further information be received in the future, a follow up report will be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL SUTURELESS AORTIC HEART VALVE
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer (Section G)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer Contact
roberta lamberti
5005 north fraser way
burnaby, bc 
MDR Report Key15287295
Report Number3004478276-2022-00179
Device Sequence Number2406990
Product Code LWR
UDI-Device Identifier00896208000443
UDI-Public(01)00896208000443(240)ICV1211(17)230923
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/23/2023
Device Model NumberPVS27
Device Catalogue NumberICV1211
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer08/29/2022
02/24/2026
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date09/26/2022
02/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight61 KG
-
-