• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This incident has been recorded under b)(4).Once investigation is completed a final/supplemental report will be submitted.
 
Event or Problem Description
It was reported that outside of surgery it was noted that the power inlet module's fuse box was burnt.No harm or delay were reported as it relates to this event.No adverse events were reported.
 
Additional Manufacturer Narrative
This incident has been reported under (b)(4).Review of the most recent repair record determined that the power inlet module was burnt.The power inlet module was replaced and resolved the reported issue.The root cause of the reported issue is attributed to the time the high flow valve is left open is too short during the pump start up sequence.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR
Common Device Name
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15287345
Report Number0001954182-2022-00084
Device Sequence Number2278938
Product Code JCX
UDI-Device Identifier00889024466005
UDI-Public(01)00889024466005
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K190789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00514010200
Device Lot Number0042975
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-