Model Number 419688 |
Device Problems
Failure to Capture (1081); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Dyspnea (1816); Pleural Effusion (2010); Pulmonary Edema (2020); Obstruction/Occlusion (2422); Heart Failure/Congestive Heart Failure (4446); Swelling/ Edema (4577)
|
Event Date 03/17/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: 694765 lead implanted: (b)(6) 2011; icf09b45 lead implanted: (b)(6) 2004.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient presented to the emergency department with shortness of breath and lower extremity edema.Chest x-ray showed pulmonary vascular congestion, mild interstitial edema, mild cardiac enlargement and a small pleural effusion.A venogram revealed 1.5 centimeter occlusion of the left subclavian vein with collaterals filling the remaining brachiocephalic vein and superior vena cava.The patient was hospitalized for congestive heart failure and administered diuretic medication.The left ventricular (lv) lead was capped and replaced with a his bundle lead.No further patient complications have been reported as a result of this event.The patient is a participant in a clinical study.
|
|
Event Description
|
Additional information received indicated it was further reported that the lv exhibited a loss of capture.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|