| Model Number |
100/189/070 |
| Medical Device Problem Codes |
Deflation Problem (1149); Pressure Problem (3012)
|
| Health Effect - Clinical Code |
Pneumonia (2011)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other, other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that there was cuff deflation and malfunctioning of a pressure valve.Ventilator associated pneumonia was reported.Additional information was requested; however, no further information was received.
|
| |
|
Additional Manufacturer Narrative
|
|
Two photos were received.Based on the photos the reported failure mode could not be detected.The complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No action has been taken since the complaint was not confirmed.
|
| |
|