Model Number LF-410540
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Failure of Implant (1924); Seroma (2069)
Date of Event 07/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported right side rupture and exchange from texture to smooth implant.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, one opening on the posterior side assessed as surgical impact opening.(shell thickness was within specification) anxiety-product/procedure: unable to observe since it is a medical event and is not related to the device.Additional observations: crease fold observed on the device.Wear abrasion observed on the device.No penetrating nick (stress marks) observed on the device.No further actions are required as the device was implanted.
 
Event or Problem Description
Healthcare professional reported right side rupture and exchange from texture to smooth implant.Device has been explanted.
 
Additional Manufacturer Narrative
The reported events are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.
 
Event or Problem Description
Healthcare additionally reported seroma and capsular contracture baker grade unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
2525 dupont drive
building 3
irvine, CA 92612
8479366324
MDR Report Key15288369
Report Number9617229-2022-14336
Device Sequence Number2504562
Product Code FTR
UDI-Device Identifier10888628003156
UDI-Public10888628003156
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/17/2019
Device Model NumberLF-410540
Device Catalogue NumberLF-410540
Device Lot Number2694065
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer10/14/2022
07/29/2024
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date11/10/2022
08/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient Weight69 KG
-
-