Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Procode is krd/hcg.[conclusion]: the healthcare professional reported that during a coil embolization procedure targeting an internal carotid posterior communicating artery (icpc) aneurysm, the complaint coil, a 3mm x 6cm galaxy g3 xsft (glx120306 / 30550290) was used, but there was resistance and the coil got stuck inside the concomitant excelsior® sl-10® microcatheter (stryker).The complaint coil was pushed with force, but it was able to advance through the microcatheter.A continuous flush was reportedly maintained through the microcatheter.The complaint coil was re-sheathed and the physician noticed that the embolic coil component had unraveled.It was removed together with the microcatheter and the procedure was restarted.There was no report of any negative patient impact due to the reported issue.The complaint device is not available to be returned for evaluation.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30550290) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Coil unraveling / stretching is a known potential issue associated with the use of this device.The instructions for use (ifu) provides proper handling for the device to prevent such issue from occurring.The embolic coil can become unraveled or stretched during attempts to withdraw it against resistance.With the information provided but without the complaint coil and the concomitant microcatheter available for analyses, the reported issue could not be confirmed through functional evaluation.The exact cause of the reported issue cannot be conclusively determined.Assignment of root cause for the event remains speculative and inconclusive, based on the information provided and without the return of the complaint device.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during a coil embolization procedure targeting an internal carotid posterior communicating artery (icpc) aneurysm, the complaint coil, a 3mm x 6cm galaxy g3 xsft (glx120306 / 30550290) was used, but there was resistance and the coil got stuck inside the concomitant excelsior® sl-10® microcatheter (stryker).The complaint coil was pushed with force, but it was able to advance through the microcatheter.A continuous flush was reportedly maintained through the microcatheter.The complaint coil was re-sheathed and the physician noticed that the embolic coil component had unraveled.It was removed together with the microcatheter and the procedure was restarted.There was no report of any negative patient impact due to the reported issue.The complaint device is not available to be returned for evaluation.
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on (b)(6).2022.[additional information]: on (b)(6).2022, additional information was received.On (b)(6).2022, additional information was received.The information indicated that a continuous flush had been maintained through the microcatheter regarding whether the coil had been withdrawn and repositioned, the response indicated that the coil was withdrawn with the microcatheter then only the microcatheter was re-inserted and repositioned before a new coil was inserted and implanted.There was no resistance between the guidewire and the microcatheter.The microcatheter was not repositioned over the coil while the coil was deployed or partially deployed out of the distal end of the microcatheter.When the coil was removed, there was no additional damage noted on it aside from the reported stretched / unraveled condition.The coil was purposely detached.There was no blood flow restriction / reduction as a result of the reported event.No clinically significant delay in the procedure was reported.(b)(6).The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|