ETHICON INC. MCRYL VIO 18IN 4-0 S/A P-3 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number Y464G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device analysis: visual analysis of the returned product revealed that an empty opened foil, a needle in a labeled winding former and several suture pieces of product code y464g were received to ethicon for evaluation.The suture has begun with process degradation caused by exposure to the environment and the needle detached from the strand.The product code y464g contains an absorbable suture.As the sample was received open, the time of exposure to the environment could not be determined and functional tests cannot be performed.The foil packet was visually inspected, and excessive wrinkles, delamination, and holes in the cavity were observed.This condition contributed to the degradation of the suture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this lot/batch.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
 
Event or Problem Description
It was reported that a patient underwent a quadrantectomy and suture was used.During the procedure, it was found that the needle was detached from the suture before use.It was not a control release needle.Upon evaluation of the device, the suture has begun with process degradation caused by exposure to the environment and the needle detached from the strand.As the sample was received open, the time of exposure to the environment could not be determined and functional tests cannot be performed.The foil packet was visually inspected, and excessive wrinkles, delamination, and holes in the cavity were observed.This condition contributed to the degradation of the suture.No adverse patient consequences were reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL VIO 18IN 4-0 S/A P-3 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15289267
Report Number2210968-2022-06904
Device Sequence Number12554443
Product Code GAN
UDI-Device Identifier10705031059269
UDI-Public10705031059269
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K960653
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model NumberY464G
Device Catalogue NumberY464G
Device Lot NumberLHM085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2022
Initial Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-