Model Number N/A
Medical Device Problem Codes Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the patient that she experienced a shock while treating with the bone healing system.The patient stated that the sflx 2 coil is giving her a shock when she wears it.The patient rated the pain as an 8 on a scale of 1 to 10, which is considered severe.There are no exposed nor cracked wires on the coil.The device was not dropped or in contact with any liquid.The patient advised that her ankle is still tender.The patient did not speak to a doctor regarding the issue.The patient will return the old coil.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Date of event: the event occurred sometime in (b)(6) 2022.Medical product: unknown.Therapy date: unknown.Customer has indicated that the product is in process of being returned to zimvie for investigation.Once the investigation is complete, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The device was not returned to highridge medical for evaluation.The reported event was unable to be confirmed due to limited information received from the customer.The device history record was reviewed, and no discrepancies related to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Highridge medical will continue to monitor for trends.
 
Event or Problem Description
It was reported by the patient that she experienced a shock while treating with the bone healing system.The patient stated that the sflx 2 coil is giving her a shock when she wears it.The patient rated the pain as an 8 on a scale of 1 to 10, which is considered severe.There are no exposed nor cracked wires on the coil.The device was not dropped or in contact with any liquid.The patient advised that her ankle is still tender.The patient did not speak to a doctor regarding the issue.The patient will return the old coil.No additional patient consequences have been reported.No product was returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFLX 2 COIL
Common Device Name
COIL
Manufacturer (Section D)
EBI, LLC.
1 gatehall dr
suite 303
parsippany NJ 07054
Manufacturer (Section G)
EBI, LLC
1 gatehall dr
suite 303
parsippany NJ 07054
Manufacturer Contact
brianne mcgill
1 gatehall dr
suite 303
parsippany, NJ 07054
9732999300
MDR Report Key15289332
Report Number0002242816-2022-00099
Device Sequence Number17968871
Product Code LOF
UDI-Device Identifier00812301020515
UDI-Public00812301020515
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P790002/S026
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberN/A
Device Catalogue Number1068226
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/24/2022
Supplement Report FDA Received Date06/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-