| Model Number |
1581-22-115 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Fall (1848); Pain (1994); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577)
|
| Date of Event |
08/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient presented with pain and swelling secondary to a fall on (b)(6) 2022.
|
| |
|
Event or Problem Description
|
|
Medical records received: clinic note dated 04 august 2022: patient presented with pain and swelling secondary to a fall four months earlier on (b)(6) 2022.Patient reports walking difficulty requiring the use of a cane and a pain level of 5.The knee was aspirated for a pre-operative synovasure test, which returned a positive result.No additional records are included in the note.The patient received a left knee revision to treat pain on (b)(6) 2022.No confirmation of infection.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.All available photo evidence were reviewed and does not shows an implant fracture, disassociation or any indicative related to a device nonconformance.Based on the information provided as evidence, a defect or malfunction cannot be established, therefore the allegation is not confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|