Model Number MAF-GM
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been over 6 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, it is likely the coating peeled off due to a usage stress, external factors, or handling.However, a definitive root cause could not be determined.During inspection and testing, the reported issue (pinhole on bending section cover) was confirmed and found to be caused by external factors.It was observed there a leakage due to a pinhole.In addition, service found the insertion tube was crushed due to external factors.There was a leakage from the digital camera, and the cross/ok/menu buttons due to external factors.Per the legal manufacturer, there is no potential for these issues to cause or contribute to death or serious injury if the malfunctions were to recur.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer reported there was a pinhole on the bending section cover of the subject device.There was no patient or user injury reported due to the event.The device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, it was observed the coating on the insertion tube was peeling off.This report is being submitted for the malfunction found during evaluation (coating peeling off).Additional details were requested regarding the reported event; however, no further information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRWAY MOBILESCOPE
Common Device Name
AIRWAY MOBILESCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15290134
Report Number9610595-2022-01279
Device Sequence Number16717324
Product Code EOQ
UDI-Device Identifier04953170340369
UDI-Public04953170340369
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082720
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAF-GM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-