Model Number 24703
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 2.5mm stenosed target lesion was located in the non- tortuous and mildly calcified femoral vein.A 3.0mm x 150mm x 150cm sterling sl balloon catheter was advanced for dilation.The physician attempted to inflate the balloon twice at 8 atmospheres (atm) but failed, the device was then removed and re-entered into the lesion site.The balloon was partially inflated at 8 atm.So, the physician increased the atmospheres to 10 atm, the balloon was fully deployed after 5-10 minutes.The device was removed and was introduced into another lesion to inflate again.However, during inflation at 8 atmospheres, the balloon ruptured.The device was simply pulled out and the procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 2.5mm stenosed target lesion was located in the non- tortuous and mildly calcified femoral vein.A 3.0mm x 150mm x 150cm sterling sl balloon catheter was advanced for dilation.The physician attempted to inflate the balloon twice at 8 atmospheres (atm) but failed, the device was then removed and re-entered into the lesion site.The balloon was partially inflated at 8 atm.So, the physician increased the atmospheres to 10 atm, the balloon was fully deployed after 5-10 minutes.The device was removed and was introduced into another lesion to inflate again.However, during inflation at 8 atmospheres, the balloon ruptured.The device was simply pulled out and the procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling sl balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the guidewire lumen is prolapsed 4.2cm from the tip.Microscopic examination revealed a hole in the guidewire lumen at the prolapse.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a hole in the guidewire lumen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING SL
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15291024
Report Number2124215-2022-32315
Device Sequence Number11335735
Product Code DQY
UDI-Device Identifier08714729782315
UDI-Public08714729782315
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093720
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/10/2024
Device Model Number24703
Device Catalogue Number24703
Device Lot Number0028843074
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/15/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
Patient Weight62 KG
-
-