| Model Number |
24703 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 2.5mm stenosed target lesion was located in the non- tortuous and mildly calcified femoral vein.A 3.0mm x 150mm x 150cm sterling sl balloon catheter was advanced for dilation.The physician attempted to inflate the balloon twice at 8 atmospheres (atm) but failed, the device was then removed and re-entered into the lesion site.The balloon was partially inflated at 8 atm.So, the physician increased the atmospheres to 10 atm, the balloon was fully deployed after 5-10 minutes.The device was removed and was introduced into another lesion to inflate again.However, during inflation at 8 atmospheres, the balloon ruptured.The device was simply pulled out and the procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 2.5mm stenosed target lesion was located in the non- tortuous and mildly calcified femoral vein.A 3.0mm x 150mm x 150cm sterling sl balloon catheter was advanced for dilation.The physician attempted to inflate the balloon twice at 8 atmospheres (atm) but failed, the device was then removed and re-entered into the lesion site.The balloon was partially inflated at 8 atm.So, the physician increased the atmospheres to 10 atm, the balloon was fully deployed after 5-10 minutes.The device was removed and was introduced into another lesion to inflate again.However, during inflation at 8 atmospheres, the balloon ruptured.The device was simply pulled out and the procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling sl balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the guidewire lumen is prolapsed 4.2cm from the tip.Microscopic examination revealed a hole in the guidewire lumen at the prolapse.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a hole in the guidewire lumen.
|
| |
|
Search Alerts/Recalls
|