• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 7MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 7MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035477200
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the basilar artery aneurysm procedure, the subject coil was broken while advancing within the microcatheter prior to the deployment.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes), there are controls in the manufacturing process to ensure the product met specifications upon release.During visual/microscopic inspection, the coil delivery wire kinked/bent.The main coil was found to be severely stretched.The main coil suture was found to be damaged.There was no coil breakage noted.During functional inspection, main coil broken/fractured during use functional test was not applicable.The defect was not confirmed during visual inspection.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was not confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.During analysis, the main coil was found to be severely stretched, the main coil suture was found to be damaged, and the coil delivery wire was found to be kinked/bent.An assignable cause of not confirmed will be assigned to the as reported ¿main coil broken/fractured during use¿ as the defect was not confirmed during the analysis.An assignable cause of procedural factors will be assigned to as analyzed ¿main coil stretched¿, ¿main coil suture damage¿ as these defects are associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural and/or anatomical factors during use.An assignable cause of handling damage will be assigned to the as analyzed 'coil delivery wire kinked/bent' since it is most likely that this damage occurred due to handling of delivery wire during the clinical procedure.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Event or Problem Description
It was reported that during the basilar artery aneurysm procedure, the subject coil was broken while advancing within the microcatheter prior to the deployment.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 7MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15291337
Report Number3008881809-2022-00410
Device Sequence Number20624698
Product Code HCG
UDI-Device Identifier04546540676573
UDI-Public04546540676573
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2024
Device Model NumberM0035477200
Device Catalogue NumberM0035477200
Device Lot Number22902527
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECHELON 10 MICROCATHETER (MEDTRONIC); NEUROFORM ATLAS STENT SYSTEM (STRYKER)
Patient SexUnknown
-
-