• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR FOOT PEDAL; REUSABLE BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR FOOT PEDAL; REUSABLE BIOPSY Back to Search Results
Model Number FSENCOR01
Medical Device Problem Codes Failure to Read Input Signal (1581); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the serial number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 08/2045).
 
Event or Problem Description
It was reported that during a breast biopsy procedure, the device allegedly took samples from the wrong quadrant.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a device history record and manufacturing review was not required as the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the encor fire forward device was received at the bd-bard global service & repair.The encor fire forward was visually inspected upon receipt and was found to be in good overall condition and functions normally.Further, found that the silver serial number label has a gouge in it exposing the painted metal underneath, there are paint chips on the edges and along the bottom exposing bare metal, and the serial number label upgrade need to be performed.The device was functionally tested while manipulating the cord and passed the test as all three switches responded normally and this was tested ten times with each switch.No other anomalies were identified.Therefore, the investigation is determined to be unconfirmed for the reported device is taking samples from the wrong quadrant and the device is not responding.The root cause for the reported device is taking samples from the wrong quadrant cannot be determined as the problem could not be reproduced.The root cause for the reported device is not responding also cannot be determined as the problem could not be reproduced.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 08/2045), g3, h6 (device).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a breast biopsy procedure, the device allegedly took samples from the wrong quadrant.Further it was reported that the device was not responding intermittently.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR FOOT PEDAL
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15291369
Report Number2020394-2022-00682
Device Sequence Number12547500
Product Code KNW
UDI-Device Identifier00801741086557
UDI-Public(01)00801741086557
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K013641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFSENCOR01
Device Catalogue NumberFSENCOR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer10/15/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-