| Model Number |
10430603 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The cse (customer service engineer) found a defective internal network switch and replaced it during on-site service activities.After the network switch was replaced, the system again worked as specified.The occurrence rate of the error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.Based on the information available, no general product design issue could be identified, and the issue is rated as a service case.
|
| |
|
Event or Problem Description
|
|
It was reported to siemens that a malfunction occurred while operating the somatom definition flash ct system.On (b)(6), 2022 a 6 month old child was planned for a procedure when the gantry unexpectedly shut down.According to the available information, the communication between the image reconstruction system (irs) and the image control system (ics) was disturbed due to a network error.No negative health consequences were reported for the child in association with the event and a rescan was not conducted.This report has been submitted with an abundance of caution.
|
| |
|
Search Alerts/Recalls
|