Model Number 10430603
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The cse (customer service engineer) found a defective internal network switch and replaced it during on-site service activities.After the network switch was replaced, the system again worked as specified.The occurrence rate of the error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.Based on the information available, no general product design issue could be identified, and the issue is rated as a service case.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the somatom definition flash ct system.On (b)(6), 2022 a 6 month old child was planned for a procedure when the gantry unexpectedly shut down.According to the available information, the communication between the image reconstruction system (irs) and the image control system (ics) was disturbed due to a network error.No negative health consequences were reported for the child in association with the event and a rescan was not conducted.This report has been submitted with an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DEFINITION FLASH
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key15292443
Report Number3004977335-2022-31782
Device Sequence Number17905961
Product Code JAK
UDI-Device Identifier04056869006963
UDI-Public04056869006963
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Remedial Action Inspection,Repair
Type of Report Initial
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10430603
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2022
Initial Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age6 MO
Patient SexUnknown
-
-