MAUDE Adverse Event Report: CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 18FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 18FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884720189
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported there was a hole in the balloon.During use they realized the balloon was emptying.The feeding tube was replaced and the second feeding tube also had a hole in the balloon and was replaced.There was no injury to the patient.
 
Additional Manufacturer Narrative
Additional information: h4 manufacture date was added.The device history record (dhr) review was performed showing no discrepancy was found according to the failure mode reported.A sample analysis could not be performed because no photo or sample was available for evaluation.The complaint will be reopened if a sample is received.The reported condition could not be confirmed.A root cause could not be determined with the available information.The manufacturing process was reviewed.Process and controls were found properly followed, including sub-assemblies, finished product assembly, packaging and inspections performed to the product.There were no abnormal conditions found that could trigger the reported condition.The current process is running according to product specifications meeting quality acceptance criteria.As part of the containment, a notification was made to the production personnel of the line so that they are aware of the condition reported by the client.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 18FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15293183
Report Number9612030-2022-03364
Device Sequence Number7054572
Product Code KNT
UDI-Device Identifier10884521007284
UDI-Public10884521007284
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720189
Device Catalogue Number8884720189
Device Lot Number2106834864
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date09/29/2022
Date Device Manufactured03/19/2021
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-