ABBOTT MEDICAL PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problem Battery Problem (2885)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/01/2022
Event Type  Injury  
Event Description
It was reported the patient's ipg is inoperable.Surgical intervention may take place at a later date.
 
Manufacturer Narrative
Date of event is estimated.
 
Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete patient information.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information received indicates the ipg was explanted and replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15293509
MDR Text Key298587858
Report Number1627487-2022-04646
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/27/2018
Device Model Number3660
Device Catalogue Number3660
Device Lot Number5363117
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS ANCHORS X 2; SCS LEADS X 2
Patient Outcome(s) Other;
Patient SexFemale
-
-