Model Number 80618
Medical Device Problem Codes Break (1069); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a break occurred resulting in additional intervention.A direxion catheter was selected for use to treat an endoleak in the lumbar artery.During withdrawal, the catheter was difficult to remove and ripped.It took one and a half hours to retrieve the broken part of the catheter from the patient.The patient may come back and get treated again.There were no patient complications.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the direxion microcatheter was returned to boston scientific for analysis.The device shaft was microscopically analyzed for any damage.The device showed multiple fractures and bends located on the shaft.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
 
Event or Problem Description
It was reported that a break occurred resulting in additional intervention.A direxion catheter was selected for use to treat an endoleak in the lumbar artery.During withdrawal, the catheter was difficult to remove and ripped.It took one and a half hours to retrieve the broken part of the catheter from the patient.The patient may come back and get treated again.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15293876
Report Number2124215-2022-31840
Device Sequence Number12321405
Product Code KRA
UDI-Device Identifier08714729839354
UDI-Public08714729839354
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/14/2023
Device Model Number80618
Device Catalogue Number80618
Device Lot Number0028553496
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient SexMale
-
-