MAUDE Adverse Event Report: COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number G49945
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that when a universa soft ureteral stent set was opened, the user noted the stent had a tear on the untethered end.The tether was attached to the stent when the set was opened.The user completed the procedure with a new device of the same product reference.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Correction: b1, h1- this report is being sent to indicate the complaint event is not reportable under fda 21 cfr part 803.As such, this event no longer meets the set criteria for a reportable event; no further reports regarding this event will be submitted.Based on the investigation, including a physical examination of the returned device, cook has determined that the stent was torn by the tether rather than a tear of the untethered device end, which our initial report was based on.There is no evidence to suggest that the failure mode stent torn by tether would be likely to cause or contribute to a death or a serious injury if the failure were to recur.Furthermore, there is no evidence that the device caused or contributed to a serious injury or patient death.As such, the event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury, or reportable product malfunction.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA SOFT URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15294227
Report Number1820334-2022-01405
Device Sequence Number2658723
Product Code FAD
UDI-Device Identifier00827002499454
UDI-Public(01)00827002499454(17)250120(10)14486950
Combination Product (Y/N)N
PMA/510(K) Number
K151051
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2025
Device Model NumberG49945
Device Catalogue NumberUSH-600
Device Lot Number14486950
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/15/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-