| Model Number |
DBP-200SOLID145 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
08/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
|
| |
|
Event or Problem Description
|
|
After a treatment in a 6 mm, totally occluded lesion from the mid to distal superficial femoral artery (sfa) on low speed, medium speed, and high speed, using a diamondback 360 peripheral orbital atherectomy device (oad) the crown stopped spinning.The crown was temporarily stuck.Using the advancer knob, the device was retracted and removed from the body.A perforation in the sfa was noticed.The perforation was resolved with balloon angioplasty and stent placement.Following the procedure, the patient was responsive with good flow to the foot.
|
| |
|
Additional Manufacturer Narrative
|
|
The oad was received at csi for analysis.Analysis revealed adhered biological material on the driveshaft crown section.The material caused resistance when passing a wire through.Analysis did not identify any damage that may have contributed to the accumulating material.The morphology and exact root cause of the accumulated biological material is unknown.When tested the device functioned as intended.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.H6 code 4247: suggested code is biological material present on device.Csi id: (b)(6).
|
| |
|
Search Alerts/Recalls
|