| Catalog Number |
05021537001 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The meter was requested for return.The meter has not been received at this time.Investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they had a display issue with a coaguchek xs plus meter serial number (b)(4).The meter display has faded segements.Meter displays the roche logo when powered on but the rest of the display has black and white squiggles on it.No misinterpretation of results was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The meter was received for investigation.Several screens were checked and the complete display works properly.The device was disassembled for inspection and no contamination or other abnormalities were visible.The reported issue could not be reproduced.The meter is functioning as intended.Medwatch fields d9 and h3 have been updated.
|
| |
|