MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number DAT015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Discomfort (2330)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A practice administrator reported that following an intraocular lens (iol) implant procedure, the patient experienced temporal arc, poor vision, and visual discomfort.The lens was explanted in a secondary procedure after 6 months of initial implantation.The clinical reasons for the explant were dysphotopsia, intolerable to optics and inability to adapt.The patient received a replacement monofocal lens.Additional information was received stating, in the surgeon's opinion, a quality issue with the iol did cause or contribute to the event indicating patient could not see well, waxy vision, quality issue.It was also indicated that extended depth of focus (edof) lens was not compatible for patient.There was no patient harm.
 
Additional Manufacturer Narrative
Information was provided that indicated the use of a qualified cartridge and viscoelastic.The product investigation could not identify a root cause for the reported complaint.The explanted lens was not returned for an evaluation.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was received indicating that patient received a replacement lens (not company).The lens was an +20.5d diopter.No patient harm was reported.Explanted lens is not available for return.In the surgeon's opinion, a quality issue with the iol did cause or contribute to the event indicating ¿patient was 20/20, j1, but felt she could not see well, waxy vision, quality issue.¿ it was also indicated that ¿edof lens was not compatible.Based on information that the company 21.0 (1.5 extended depth of focus) lens was exchanged for a non-company monofocal one half diopter in strength suggests that the replacement lens was an improved selection for this patient's vision needs or preferences.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15294967
Report Number1119421-2022-01852
Device Sequence Number17724898
Product Code POE
UDI-Device Identifier00380652424653
UDI-Public00380652424653
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/21/2026
Device Model NumberDAT015
Device Lot Number15221643
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer10/14/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION HANDPIECE; DISCOVISC; MONARCH III CARTRIDGE D; ORA WITH VERIFEYE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-