• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-80/I20-40
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft, a suprarenal stent graft extension and an ovation extender to treat an abdominal aortic aneurysm (aaa).Approximately 4 years post initial procedure during a routine follow up a type 3a endoleak was identified.Patient is pending a follow up angiogram and the physician is planning a re-intervention.
 
Additional Manufacturer Narrative
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type 3a endoleak (of the bifurcated stent graft and proximal aortic extension) with additional endovascular procedure are confirmed.This is consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.While the initial procedure was off-label due to concomitant device usage, it is unclear if this contributed to the reported event.The final patient status was reported as doing well.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.
 
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft, a suprarenal stent graft extension and an ovation extender to treat an abdominal aortic aneurysm (aaa).Approximately 4 years post initial procedure during a routine follow up a type 3a endoleak was identified.Patient is pending a follow up angiogram and the physician is planning a re-intervention.Clarification: the initial implant procedure was outside the indications of use (off-label) due to adjunctive devices not compatible with the afx system per device ifu.Additional information: the physician elected to treat the patient by implanting one (1) afx2 bifurcated stent graft, one (1) afx vela infrarenal and one (1) afx vela suprarenal on 01 september 2022.The endoleak was successfully resolved and the patient tolerated the procedure well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15296028
Report Number2031527-2022-00213
Device Sequence Number14117995
Product Code MIH
UDI-Device Identifier00818009014412
UDI-Public(01)00818009014412(17)190115
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other,Recall
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/15/2019
Device Model NumberBEA28-80/I20-40
Device Lot Number1776266001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer10/14/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/08/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL LOT#1737146003; OVATION IX EXTENDER LOT#FS032118-02
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age74 YR
Patient SexMale
-
-