WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEEDSHIFT COMP IMPL KIT 20X20X20 OFST10; STAPLE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
SE-2020-10 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the product code (b)(4), the wrong drill guide was packaged with the product.The 15-06 drill guide, to be use with (b)(4), was packaged with the product.There were no surgical delay.Procedure was completed successfully.There was no patient consequences.This report is for one (1) speedshift comp impl kit 20x20x20 ofst10.This is report 1 of 1 for (b)(4).Complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: h3, h4, h6: part se-2020-10, bme lot mse21023: manufacturing date or release to warehouse date: march 03, 2022.Place of manufacture: biomedical enterprises, san antonio, tx.Lot expiration date: november 01, 2026.No non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.This lot met all dimensional, visual, and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.H3, h6: a photo investigation was completed: visual analysis of the photo was performed.It was confirmed that the 15-06 drill guide, to be use with the product code se-1520-06, was packaged with the product se-2020-10.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed.The root cause of the complaint condition can be confirmed due to the manufacturing issue.The received condition was confirmed.Relevant actions have been taken to address the issue.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|